ENERGY
SOURCES IN SURGERY
Dr.
M. Ramesh, D.N.B, F.I.C.S.
Introduction
Surgery
today is being driven by technological innovations more
than ever before. We are also becoming dependent on technology
for further progress. I would not be wrong when I say
that a whole new chapter of surgery was opened with the
invention of Electro surgery several decades ago.
Adding
better instrumentation and gadgets to our armamentarium
is certainly helping us to leap ahead in the way we perform
and our results are being made “better-than-before” always.
Electro surgery not only helped us to perform surgery
faster but it has also made us do certain operations that
we had not imagined to be possible. Understanding nature
and the energies around us, and the right application
of them has surely been the prime factor to spell success
in the story of the surgeon.
Today there is hardly any surgery where energy sources
are not used. Yet it remains a hard truth that not a single
surgical textbook that we read during our student days
had a dedicated chapter for this most important factor
that can make or break the fate of the patient. It has
always been an enigma to me as to why we are not taught
about these energy sources in our surgical curriculum.
There is no doubt that the clear understanding of these
energy sources is most essential for our surgical practice.
I
have made a sincere attempt to write about the different
gadgets that we use every day and what I feel is absolutely
essential for every surgeon to know. This is a concise
compilation of data rather than an article based on a
surgeon’s experience.
Electro
surgery
This is to describe the basic principles of Electro surgery.
Unlike lasers, there has not been any regulatory body
on the use of electro surgery. Traditionally the use of
electro surgery has been learnt from the seniors during
surgery, and surprisingly there is hardly anything written
about it in any of our standard textbooks of surgery.
Inspite of its potential dangers, there is not enough
effort to understand the occurrence and prevention of
the complications of electro surgery. This article attempts
to give an insight into this topic of utmost importance
to any surgeon today.
Definition
One common mistake that we often come across is that the
terms “Electro surgery” and “Cautery” is used for the
same purpose while both of them are quite different. By
definition, Electro surgery is the use of radio frequency
alternating current to raise the cellular temperature
as a way to vaporize or coagulate tissue. Cautery is a
term derived from “Kauterion” which means, “hot iron”.
It is the destruction or denaturation of tissue by a passive
transfer of heat or application of a caustic substance.
Biological
Effects of Electricity
There are primarily three types of effects produced on
tissues by electrical energy. The first type is called
Electrolytic effect where anions and cations in the body
are attracted to opposite sides. This type of reaction
is not conducive to life, and is produced by either low
frequency alternating current or direct current. The second
type of reaction is called Faradic effect, which is produced
by high frequency alternating current up to 20 KHz. This
type of current causes stimulation of nerve-endings and
muscles, and is commonly used by physiotherapists and
neurologists, and sometimes by general surgeons during
parotid surgery to detect the branches of the facial nerve,
etc. The third is the Thermal effect, which is produced
by high frequency alternating current more than 300 KHz
(also called radio-frequency AC).
There
are basically three variable properties of electricity
- Current, Voltage and Resistance. Current (I) is a measure
of the electron movement past a given point in the circuit
in a fixed period of time. It is measured as amperes.
Voltage (V) is the pressure with which the electrons are
pushed through the tissue. This is measured as volts.
Resistance (R) is the measure of the difficulty that a
given tissue presents to the passage of electrons, and
is measured as ohms. Power (W) is the capacity to do work
per unit time and is measured in watts. All this can be
very easily understood using Odell’s Water-tower analogy.
Electro
surgical Unit (ESU)
An
Electro surgical Unit basically does two major functions.
It converts a 60 cycles/second (60 Hz), low voltage alternating
current into higher voltage radio frequency (500 KHz to
3.0 MHz) current. Secondly it is capable of producing
current with a variety of waveforms.
Advantages
of Electrocutting
u Reduced bleeding as there is simultaneous haemostatic
effect. Preclusion of germ implantation, as there is heat
produced in the vicinity, and as it is done by sterile
technique.
u
Avoidance of mechanical damage to the tissue.
u
The possibility of using it in endoscopic surgery.
Types
of Circuits
There are two types of circuits used to produce diathermy,
Monopolar and Bipolar. In the Monopolar diathermy, the
electricity travels from the ESU to the patient. The current
enters the body of the patient and reaches the dispersive
electrode (patient plate), which may be at a distance
from the active electrode, and then returns the ESU. As,
alternating current is used, the direction of current
keeps changing several times every second. In the bipolar
diathermy, the current passes through one limb of the
instrument and returns through the other limb of the instrument.
While doing so, it travels through the tissue grasped
between the two limbs of the instrument.
In
the Monopolar type the effect of the current is seen at
close proximity to the active electrode, as the energy
is concentrated here, and gets dispersed as it travels
towards the dispersive or return electrode. The advantages
of using the Monopolar Electrocautery are that it is easy
to use and surgery can be performed much faster as it
can be used as both cutting and coagulating current and
thus can be used to dissect tissues also. The disadvantage
is that larger volumes of tissue are injured and sometimes,
distant burns can also occur. It requires a distant return
electrode. It may also interfere with pacemakers. To prevent
complications it is important to place the return electrode
(patient plate) as close to the operating field as possible,
so that the circuit runs only for a short distance in
the patient’s body. Generally, it is advised to place
the dispersive electrode around the arms or the thighs
depending in which part of the body the surgery is being
performed. The advantage of using Bipolar Electrocautery
is that small volumes of the tissue are injured and there
will not be any distant burns. It is a safe mode when
used in patients with pacemakers. It is also effective
in wet fields. The main disadvantage is that more skill
and time is required to use bipolar Electrocautery, and
that only coagulation current is available. Hence there
is no dissecting capability. But some of the recent machines
have incorporated the cutting mode also. Bipolar offers
more safety when being used at close proximity to bowel
and other abdominal viscera.
Tissue
Effects of Electro surgery
There are three types of tissue effects of the radio frequency
current, which is used for electro surgery - Vaporization
(or Cutting), Desiccation (or Coagulation) and Fulguration
(which is superficial coagulation). Vaporization and fulguration
are non- contact procedures, and there is a small distance
between the electrode and the tissue. The electrical spark
travels through a steam bubble from the tip of the active
electrode to the tissue to cause the particular effects
on the tissue.
A high-frequency current with a sinusoidal waveform results
in a pure cutting effect. The coagulation mode is produced
by a series of dampened sinusoidal waves that are produced
in rapid bursts. A combination of both cutting and coagulation
effects is possible with a current that is a blend of
the pure sinusoidal cutting current and the periodic dampened
coagulation current. Altering the blended waveform can
vary the relative cutting versus coagulation effect.
Mechanism
of action (Cutting / Coagulation)
When
an alternating current is used on a cell at a very high
frequency (radio frequency), the anions and cations move
to and fro within the cell with each cycle of the alternating
current. This causes friction and results in increase
in the intracellular temperature. Vaporization or cutting
is caused by high current and low voltage. This causes
a rapid heating of the cell and formation of steam inside.
As a result there is an explosion due to the massive increase
in volume of the intracellular contents, and lysis of
the cell takes place. A low current and high voltage produces
coagulation. This is a damped current and the flow of
current is interrupted. Though the increased voltage causes
deeper penetration into the tissues, the low current causes
slow heating of the cell. This in turn causes dehydration
of the cell and the cell shrinks in size. It is important
to recapitulate the Ohm’s Law, which says: I = V/R This,
when applied to the formula: W =VxI, =Vx(V/R) = V2 /R
where, I = Current, V = Voltage, R = Resistance, and W
= Power.
Hence
the amount of work done (coagulation performed) is directly
proportional to the voltage used and is indirectly proportional
to the resistance offered by the nature of tissue on which
it is used. A higher voltage leads to a higher spark intensity
and a higher spark intensity results in a deeper zone
of coagulation during the cutting process. The variables
that affect the tissue effects of R-F current are as follows:
u
Generator output
u
Power density (Size and shape of electrodes)
u
Electrode tissue proximity
u
Tissue impedance
u
Electrode speed / time on tissue
u
Distension media.
Ideally
when tissues have to be cut, a sharp electrode is used
in the cutting mode, and the electrode is held at a small
distance away from the tissue. Charring effect will be
minimal when used in this way. It must be remembered that
the electrical current generates the cutting effect of
the electro surgical unit, not by the pressure of the
electrode against the tissues/ There should be virtually
no resistance to the movement of the tip of the electrode
to the tissue. If resistance is encountered one must be
sure that the active electrode is clean, that only the
tip of the electrode is in contact with the tissue, and
the power is at an adequately high setting. When fulguration
(superficial coagulation) is desired, as while obtaining
haemostasis over the liver bed, a ball electrode or a
spatula is used in the coagulation mode, again holding
the electrode at a small distance away from the tissue.
If the electrode is pressed firmly over the liver surface,
it would cause desiccation or deep coagulation. When coagulation
or desiccation is desired, a flat electrode or a grasping
instrument is used in the coagulation mode, in full contact
with the tissue. While dissecting tissues and both cutting
and coagulation are required, blended modes are used in
different ratios of cutting and coagulation. The thickness
of the electrode can be selected depending on how much
coagulation effect is desired. The speed at which the
electrode is moved determines the amount of contact and
delivery of energy to the tissues, and thus the amount
of coagulation and charring at the margins. If the coagulation
current is used to incise tissue, tension and counter-tension
has to be applied to the tissue, otherwise separation
does not occur and it results in excessive thermal injury
to neighbouring tissue. In laparoscopy, the Carbon dioxide
gas used is not as good a conductor of the electrical
energy as air, and thus would alter the performance of
electro surgery. The power of the cautery should be set
at the lowest possible setting to produce the desired
effect of either cutting or coagulation, and prevent charring
of neighbouring tissue.
When
using the monopolar electrocautery to achieve haemostasis,
it is important to remember that the electrical current
will diffuse throughout any conductor of electricity including
blood. Therefore, is imperative that the field in general
be dry at the time of application of the current.
The
bipolar cautery seals vessels by a coaptive technique.
While using bipolar cautery, the tips of the forceps should
not come in contact with each other. This will produce
short-circuiting of the current and the tissue effects
decrease. The tissue should be lightly grasped to allow
a 1 to 2mm gap between the electrode tips. Coagulation
with the bipolar forceps tends to be a self-limiting process
since desiccation of tissue between the points will ultimately
prevent further flow of current. For efficient use of
the bipolar forceps, the tips must be free from the buildup
of charcoal and coagulum. Therefore, frequent cleansing
of the tips is required.
Electro
surgical burns
During application of elecrosurgery three main types of
burns can occur :
u
Endogenous burns
u
Exogenous burns
u
Psuedo burns
ENDOGENOUS
BURNS
Endogenous
burns are always a result from a too high current density
in the patient’s tissue. At the active electrode there
is a need for high current density in order to cut or
coagulate tissue, but accidental pressing of the foot
pedal or use of the electrocautery for a longer time or
extent can cause burns. There are three mechanisms in
which inadvertent burns can occur during use of monopolar
electrocautery in laparoscopic surgery.
Direct Coupling:
The
common cause by which this occurs is when there is insulation
failure or when the whole metal part of the instrument
is not being visualized while using elecrocautery. The
instrument may be touching some other tissues outside
the laparoscopic visual field, where the instrument may
not be insulated adequately. At such a time the current
passes to that tissue and causes burns there.
Indirect Coupling:
There
can be another instrument which can conduct electricity
(like telescope, grasper, etc.,), in close proximity to
the instrument through which electricity is passing and
the energy can jump and get transferred to this instrument
also. Supposing there is some other tissue in close proximity
to this instrument then there can be burns of this tissue,
which is located for away from the site of surgery. This
burn may go unnoticed.
Capacitive Coupling:
There is certain amount of energy that leaks on to the
reducer if it is made of metal, and usually if the reducer
is in contact with a metal canula on its outer aspect,
the energy is dissipated on the abdominal wall. In turn
the energy goes to the dispersive plate and returns to
the E.S.U. But if the reducer is not able to let out the
energy (because the outside canula is made of a non-conducting
material, like plastic), it gets accumulated in the reducer.
When a loop of intestine or some other viscera comes in
close proximity to the reducer, it suddenly discharges
all the energy to that tissue and can result in burns
there. Either using both metal reducer and canula, or
both being made of nonconducting material can prevent
this. The risk of capacitive coupling burns exists when
a combination of metal and plastic ports is used. Certain
modifications are incorporated in some new Electro surgical
units, like Electrosheild and monitoring devices which
are capable of either preventing accumulation of extra
energy in the portals or carrying back this energy to
the E.S.U.
The
usual causes of endogenous burns other than those mentioned
are as follows:
u Patient plate is too small
u
Patient plate is not covering the patient’s tissue with
its entire area. ( atleast 75% area should be in contact
)
u
Unintentional contact to other electrically conductive
parts, e.g., drip stand or metal parts of the operation
table, etc.,
Sometimes
there may be concentration of energy at the patient plate
or at areas where the patient comes into contact with
electric-conductive parts. The current density becomes
so high as to burn the patient’s tissue.
EXOGENOUS
BURNS
Exogenous
burns are caused from the heat of burning substances such
as skin-cleansing lotions, degreasants and disinfectants,
also anesthetics, which have been ignited by sparks between
the active electrode and the patient’s tissue. Note that
alcohol usually burns as invisible flames, since the operating
lamp illuminates brighter than the flame does. This way
the patient-burn can only be recognized after it has happened.
PSEUDO
BURNS
From time to time minor or major necrosis are found with
patients and are regarded as burns but without finding
any explanations or reasons of how these burns have been
caused. Endogenous burns can be excluded when the patient
did not have contact with electric-conductive parts at
the area where the necrosis is found. Exogenous burns
can also be excluded when before or during electro surgery
no flammable substances were used.
The
causes of these burns must be found out by differential
diagnosis: Necrosis caused by pressure to the patient’s
tissue: During long operative procedures pressure to the
patient’s tissue can cause necrosis, for example, during
heart surgery when the patient is hypothermic a large
tissue necrosis was found post-operatively. Pressure to
the patient’s skin caused by rubber-straps being used
to fix and attach the patient plate or by contact-clamps
being put underneath the patient can again cause necrosis.
In many cases this is erroneously diagnosed as patient-burn.
During electro surgery patient-burns can only occur when
the before mentioned facts exist. It is possible to prevent
patient-burns safely when the operating team knows and
observes the causes as well as pays attention to these
before and during electro surgery.
SAFETY
PRECAUTIONS
The
following steps should be followed carefully while using
electro surgery:
u
All connections are carefully checked before the ES unit
is put on. u The patient plate used is always one recommended
by the manufacturer.
u
The patient plate must always be applied by covering the
patient with its entire area as best as possible.
u
The conductive surface of the patient plate must always
be clean and free from corrosion.
u
If gelled patient plates are used, it is most important
that the gel is evenly applied over the entire conductive
area of the patient plate.
u
Prior to use, the patient plate must be checked for damage,
especially patient plates made of aluminum foil.
u It is important that the patient plate is applied with
the electrically conductive surface to the patient’s skin
and not with its wrong side.
u
The patient plate is applied as close to the operative
site as possible.
u
Care must be taken that no electrical conductive fluids
come between the patient’s skin and the patient plate.
u
The patient is insulated against all electrically conductive
objects by a thick, dry, electrically insulating sheet,
placed between the patient, the operating table and the
supports. The sheets must not become damp. Areas subject
to considerable secretion of sweat, body extremities lying
against the trunk or skin-to-skin contacts should be separated
by the application of a dry cloth. Drain off urine with
catheter.
u
During electro surgery always sparks exist between the
active electrode and the patient’s tissue.
Therefore
do not use flammable or explosive substances or gases
during electro surgery. If flammable or explosive substances
have been used, these must be completely removed before
activating the electro surgical unit.
A
special precaution to be taken during laparoscopic surgery
is that electro surgery should not be used whenever there
is bowel perforation. Bowel contains Methane gas, which
is released into the peritoneal cavity whenever there
is a bowel perforation, and if electro surgery is used
at such a circumstance, it may lead to an explosion.
Assuming
good surgical technique and good endoscopic instrumentation
with intact insulation, correct connection of cables and
proper placement of neutral electrode would go a long
way in making this efficient tool safe and a boon to the
surgeon especially in the era of laparoscopic surgery.
Argon
Beam Coagulation
Like
the standard electrocautery units, the argon beam coagulator
uses high-frequency oscillating current to generate coagulating
heat. The argon beam coagulator differs from standard
electrocautery in that it uses a spray of ionized argon
gas as the active electrode rather than a metallic blade.
This spray allows even, efficient, and broad application
of the coagulating current to the tissues. The argon beam
coagulator consists of a current generator, a grounding
pad, and a handheld active electrode. The device is activated
by a foot pedal that initiates the flow of pressurized
ion gas through the end of the handheld unit. Once a solid
column of gas connects the handheld active electrode to
the patient, the electrical current arcs across the argon
gas to the tissues. The type of current used with the
argon beam coagulator is almost identical to the type
used with standard electrocautery. To use the argon beam
coagulator, the handheld unit is held like a pencil with
the end directly pointed at the tissue from a distance
of 1 to 2 cm. The person holding the active electrode
also activates the foot pedal. (Tate’s rule). A jet of
argon gas is emitted from the end of the handheld unit,
completing the circuit between the argon beam coagulator
and the patient. The cautery current is delivered to the
surface of the tissue in contact with the argon stream.
To
cauterize large surface areas, the unit can be used like
a paintbrush using slow small strokes across the tissues.
The power settings on most units range from 0 to 150 watts.
As with the standard electrocautery, conduction is dependent
on many factors, including the conductivity of the tissue.
Therefore, it is recommended to start with a relatively
low power setting of 50 to 60 watts and increasing as
necessary. Because of its efficiency in coagulating large
irregular surfaces, the argon beam coagulator is ideal
for obtaining hemostasis along the cut surface of the
liver following hepatic resection. The argon beam coagulator
can also be helpful in controlling bleeding from minor
splenic trauma or other oozing surfaces. It has also been
used as a means of tumor debulking by fulgurating metastatic
ovarian carcinoma. The argon beam coagulator offers some
advantages over conventional electrocautery in that with
the argon beam coagulator there is no physical contact
between the active electrode and the tissues. This lack
of physical contact means that no adhesion of the active
electrode to the tissue occurs, which allows for improved
eschar integrity. In addition, there is no need to clean
the char from the instrument. The flow of argon gas blows
the blood and secretions away from the solid tissue to
be coagulated, allowing for effective and efficient coagulation
of tissue surfaces that are actively oozing, and less
smoke is generated than with conventional electrocautery.
The
argon beam coagulator has some disadvantages: its lack
of precision and its expenses both with the unit itself
as well as the consumable argon gas. In addition, although
the argon beam coagulator is excellent at standard coagulation
function, it has no coaptive capabilities and is very
limited as a dissecting tool. Because the argon beam coagulator
is essentially a monopolar electrocautery device, all
precautions applicable to monopolar cautery should apply
to the argon beam coagulator. Because of the risk of possible
injury, it is better to avoid the use of the argon beam
coagulator in close proximity to delicate structures such
as intestines, major vascular structures, ureters, and
bile ducts. With the argon beam coagulator’s relative
lack of precision, special precautions should be taken
to avoid current diversion through metallic instruments
or retractors. Inadvertent activation of the pedal can
result in significant injury and fire, and so the argon
beam coagulator should always be stored in a plastic holster
when not in use. Use in laparoscopic surgery has been
difficult as the abdomen is a closed cavity and there
can be a sudden rise in the intra-abdominal pressure with
flow of the argon gas into the abdomen.
Cryotherapy
Cryotherapy
is a technique of in situ tissue ablation that uses freezing
temperatures to cause cell death. Cryotherapy has been
used to treat a variety of benign and malignant lesions.
For several decades, in situ destruction of tumor by Cryotherapy
has been used for cutaneous lesions. More recently, this
mode of therapy has been applied to tumors of the head
and neck, cervix, rectum, prostate, breast, and liver.
Today sophisticated cryoprobes are available for the delivery
of extremely low temperatures by means of pressurized
liquid nitrogen. The probes come in varying sizes with
differing capabilities. The delivery systems are complex,
and their use requires personnel with special expertise
in their operation and maintenance.
Although
this technology is new and to a degree still unproven,
ultrasound-assisted cryotherapy appears to have great
promise in the treatment of liver tumors. In situ cryodestruction
of tumor is best applied to unresectable or multiple liver
metastases from colorectal cancer, where complete tumor
ablation may lead to improved long term survival.
Cryotherapy
causes tumor destruction and cell death by a combination
of several possible mechanisms. These mechanisms include
cold shock injury, reduction of cell volume by osmotic
dehydration, denaturation of vital cellular enzymes, perforation
of cell membranes by intracellular ice crystals, and destruction
of tumor microvasculature. To ensure complete cell death,
temperatures in the tissue should be lowered to below
- 35degree C, maintained in the frozen state for at least
3 minutes, and then slowly thawed. The thawing cycle is
particularly important, because too rapid or too slow
thawing will allow survival of a portion of the tumor
cells. For this reason, at least two freeze-thaw cycles
should be applied to each tumor to ensure complete cellular
destruction.
Infrared Coagulator
The infrared coagulator generates coagulation heat energy
by infrared irradiation. The infrared coagulator consists
of a transformer unit with a foot pedal switch and a handheld
wand. The wand is round metallic cylinder that generates
the infrared light that emanates through the crystal lens
as the end of the wand. The heat energy produced by the
infrared irradiation causes rapid heating of the tissues
in contact with the crystal, and these heating results
in desiccation and coagulation. The infrared coagulator
is most useful for coagulating oozing tissue surfaces
such as the cut edge of the liver following hepatic resection.
The flat crystal is pressed against the tissue in a manner
such that little or no light can escape from the end of
the wand. The foot pedal is then pressed to activate the
wand and generate the infrared irradiation. Approximately
1 to 2 seconds of exposure is usually sufficient to result
in tissue coagulation and hemostasis, which are signified
by the boiling of fluids at the edge of the crystal and
the generation of a small amount of smoke. It is important
that the wand not be pulled away from the tissue until
the infrared generation has ceased. After each application,
the end of the wand should be wiped with a moist sponge
to cool the crystal and to remove char from the tip. The
infrared coagulator is an effective device for coagulating
oozing surfaces and has the advantage of not requiring
electrical current to pass through the patient. Hence,
the infrared coagulator does not interfere with ECG monitoring
or pacemaker function.
Ultrasonic
Dissector
The ultrasonic dissector is a surgical tool that uses
high-frequency mechanical vibrations to fragment tissue.
Developed in the late 1960s, this technology was originally
applied to ophthalmic surgery, but has gained wide use
in neurosurgery, hepatobiliary surgery, and oncologic
cytoreductive surgery. The ultrasonic dissector system
consists of a rather bulky hand piece connected to a function
control console that is controlled by a standard foot
pedal. The end of the handheld unit consists of a metal
contact probe that vibrates at a frequency between 20,000
and 40,000 times per second. Because this vibration frequency
is above the audible range, it is referred to as ultrasonic.
No audible sound or electromagnetic radiation is emitted
and the vibrating tip must be in direct contact with the
tissues to bring about its effect. Transducers that rely
on piezoelectric crystals or magnetostrictive laminations
to convert electrical energy into mechanical vibrations
generate the vibrations.
The
ultrasonic dissector fragments tissue by contact with
high water content cells. The vibrations generate vapor
pockets within the cells that lead to cellular disruption
and fragmentation. While fragmenting high water content
cells, the dissector does not rapidly disrupt collagen-rich
tissue such as blood vessels and ducts. Hence, the device
can divide parenchymal tissue while leaving blood vessels
intact so that they can be individually ligated prior
to division.
The
most common general surgical application of the ultrasonic
dissector is for the division of the liver parenchyma
during hepatic resection. In addition to its use as a
dissecting tool, the ultrasonic dissector can be used
as a means of tissue ablation. It has been extensively
applied in cytoreductive surgery in the treatment of metastatic
ovarian cancer. Ovarian epithelial cancers tend to have
high water content with very little fibrous stroma; hence,
these tissues readily fragment with the ultrasonic dissector.
When the ultrasonic dissector is used for this purpose,
it is important to be aware that tumor infiltration can
involve full thickness penetration of hollow or tubular
structures such as intestine, bladder, and blood vessels.
Therefore, it is important that the surgeon be prepared
to deal with possible perforation of these structures.
With this in mind, it is always better to order for complete
mechanical bowel cleansing with prophylactic intravenous
antibiotics for patients undergoing cytoreductive surgery.
Additionally, prior to applying the ultrasonic dissector
to a tumor that is in close proximity to major vascular
structures, it is advisable to first gain proximal and
distal control of the vessel.
The
ultrasonic dissector is a convenient way of dividing solid
organ parenchyma with little blood loss. When used appropriately,
this device is relatively safe and mishaps are infrequent.
The ultrasonic dissector has some disadvantages. Its high
cost, the bulkiness of the unit itself, and it has not
been demonstrated to be consistently superior to standard
dissecting techniques. For most routine liver resections,
we can use a combination of electrocautery and finger
fracture technique for noncirrhotic livers and reserve
the ultrasonic dissector for patients with mild to moderate
cirrhosis.
Ultrasonic
knife
The
ultrasonically activated scalpel comprises of a high frequency
computer controlled generator, which converts incoming
electrical signals into mechanical vibrations at 55.5
kHz at the blade tip via a hand piece transducer. The
amplitude of motion amounts of 50-100 microns depending
on the power setting. The moving blade couples with the
tissue, resulting in breakage of protein hydrogen bonds
and thus protein coagulum forms. This coagulum seals off
blood vessels. The whole process operates at 80 degrees
C, minimizing undesirable tissue damage due to high temperatures.
The
ultrasonic knife can perform cutting and haemostasis with
minimal tissue damage, and visibility may be improved,
as there is less smoke. Energy flows in a longitudinal
direction thus limiting its lateral spread and thermal
injury. Unlike electro surgery, no electrical energy is
transferred to patients, and extra safety can be ensured.
In addition, the linear relationship between variables
such as duration of application does not plateau, making
the system more controllable as injury occurs gradually
and reproducibly over time. The range of blades available
serves both open and laparoscopic surgery. They are available
in both 5 and 10 mm. diameter for laparoscopy. In addition
to the grasping instrument, various other types are available
such as the hook or the ball coagulators.
Though
used for many surgeries, the most studied application
is in laparoscopic fundoplication, in which short gastric
vessels are cut haemostatically. Other applications include
laparoscopic colectomy, adrenalectomy and gastric surgery
besides others.
Lasers
The
term “Laser” is an acronym, which stands for Light Amplification
by the Stimulated Emission of Radiation. The word radiation
does not mean ionizing type of radiation, but refers to
a “radiant” body, i.e., one that “shines” with light energy.
Light is comprised of photons of energy. The normal tendency
of light is to scatter in all directions. Laser light
contains of photons released in an organized fashion called
stimulated emission.
Three
unique qualities of laser light that differentiate it
from regular light are its coherence, monochromaticity
and collimation. Coherence means that the wave patterns
of the light energy being emitted in a laser are orderly
and similar. The laser waves ore always precisely in phase
with one another, temporally and spatially. This property
of a laser would be of use for diagnostic and scanning
applications. The term monochromaticity means that lasers
produce pure colors of light. They are always of the same
wavelength and energy level. This is a property that is
useful for us because of the fact that different tissues
absorb various colours differently. Various chemical elements
emit characteristic colors of light, and the laser is
named after the material used. The term collimation means
that laser light travels with all its waves bound tightly
together, as a parallel beam in space. This particular
property of lasers is what allows the beam to be finely
focused to intensify its effects and is the major characteristic
allowing its surgical use.
The
surgical effects of the laser are due to localized heating
when the tissues absorb the light. As tissue begins to
heat, it blanches white as it coagulates, then shrivels
as it desiccates, and finally turns to steam and vapor
as it is vaporized above 100 degrees C. The heat-generating
effect of the lasers is used for surgical applications.
Lasers produce heat that is localized and produce desired
surgical effects with associated haemostasis. There are
five types of lasers primarily being used for surgical
applications. They are Carbon dioxide (CO2), Nd:YAG (Neodymium:
YAG), argon, Ho: YAG (Holmium: YAG) and the KTP (produced
by altering the infrared output of the Nd:YAG laser with
a KTP crystal). These lasers are used in two basic ways.
One is a noncontact method whereby the laser light is
absorbed by tissue and heat is generated. The other is
the contact method by which the laser heats special fibre
tips and this heat is in turn transferred to the tissue
by contact with the fibre.
The
factors that determine the amount of laser that is delivered
to the tissue are:
u Power (watts) with which the delivery system is operated.
u
Duration for which it was operated.
u
Power density (the size of focused spot used to intensify
the light).
u Color of the laser light.
u
Color and vascularity of the tissue.
The
surgeon can control the power setting, the duration of
application and the spot size of the beam, and thus modify
the effect of the laser on tissue. A laser beam can be
used either as a continuous mode or a pulsed mode. A pulsed
mode is a brief beam, which is delivered in fractions
of a second. This type of delivery gives a good control
to the surgeon to deliver precise doses of high power.
This type of delivery provides a longer reaction time
and produces less spread of heat damage to the distant
organs. In the absence of such a mode of delivery in the
laser machine, operating the foot pedal appropriately
to produce controlled bursts can still use a pulsed mode.
Any
type of lasers can be used to cut or vaporize by altering
the way it is used. For laparoscopic surgery we could
use a CO2 laser laparoscope, or regular fibres to deliver
argon, KTP, Ho:YAG and free beam Nd:YAG lasers. In a CO2
laser laparoscope, the laser beam first focuses to a point
and then diverges. Hence it maintains a long depth of
field and it can burn or vaporize tissues distal to the
target organ or tissue. Therefore while using this laser,
it is essential to have a backstop to prevent distal injury.
When a fibre is used the beam starts of diverge immediately
distal to the beam. Hence this beam does not stay focused
beyond an inch or two from the tip of the fibre. The advantage
of the fibres is that different effects can be accomplished
quickly with only a slight motion of the fingers holding
the fibre hand-piece. A small pinch back with the fingers
allows small blood vessels to be photocoagulated, then
a small pinch forward to bring the fibre end just over
tissue results in a cut, both at the same power settings
on the machine. The short range of effect also reduces
the need for intraabdominal backstops. The argon and KTP
are the primary fibreoptic lasers for laparoscopy besides
contact type modalities, which are mainly Nd: YAG lasers.
The
CO2 laser provides a great deal of versatility in the
“reach” it provides from the end of the laparoscope, the
number of angles where it can work, and the speed at which
it can vaporize or cut if desired. In this sense it is
probably more versatile than a fibre system, but it has
a significant longer learning curve and does not provide
the hemostasis than fibre systems provide. It also requires
a specialized laparoscope set and coupler to mate the
laser with the scope.
Fibres
terminating in some device such as a metal tip, sapphire
probe, or even an altered shape of the fibre tip generate
a significant amount of heat at this tip and are referred
to as hot tip devices. They act as very intense, precise
knives. Energy concepts such as power density do not really
apply to contact devices, since they rely on simple heat
conduction. The Nd: YAG laser is the primary one used
for hot tip devices in laparoscopic surgeries. Some of
the hot tip devices, such as rounded or chisel sapphire
tips, do actually focus some of the laser light, so that
combination effects may occur. A combination rounded tip
fibre can be backed off from the tissue to vaporize or
coagulate as free beam or touched to tissue to cut as
a hot tip.
The safety of the patient and the operating team is very
important. Trained, experienced assistants are invaluable
during laser procedures. Operation theatres must be clearly
labelled on the outside door indicating that a laser procedure
is in progress. The window in the operating room must
be covered. The type of laser must be specified on the
“danger” sign. The optical density for protective eye
wear must be appropriate for the wavelength of the laser
being used. All personnel, including the anesthesiologist
and assistants, must wear safety goggles or appropriate
protective eye wear. It is absolutely essential that flammable
substances or explosive solutions be avoided in the operating
room during laser usage. Special care must be taken when
using paper drapes. When using CO2 lasers the area surrounding
the operative field should be draped with moist laps.
The patient’s eyes must also be appropriately protected
with moist gauze or dressings.
The
production of smoke after tissue destruction with the
lasers has led not only to respiratory complications,
but has been implicated as a possible source of mutagenicity.
Approximately 75% of the solid particulate matter in laser
plume is less than 1 micron. When inhaled, this small
particulate matter is capable of traveling directly to
the distal tracheopulmonary tree and being deposited in
individual alveoli. To eliminate 99% of the generated
plume, a suction device that can mobilize 28 litres of
air/sec when held 1 cm from the origin of the laser plume
is needed. Although CO2 laser laparoscopy is associated
with more plume formation than the Argon, Nd: YAG, or
KTP lasers, adequate smoke evacuation systems are still
mandatory when using the latter three wavelengths.
There
are no prospective, blinded studies comparing laparoscopic
use of lasers with conventional electro surgery, which
have demonstrated a reduction in adhesion formation or
an improvement in pregnancy rates. Its use to date is
based on the clinical impression primarily of the surgeon.
Many surgeons advocate the use of lasers because of their
controlled depth of penetration, reduced thermal damage
to adjacent tissue, and the reproducibility of the effects
of lasers on tissue.